Intelect NMES Digital | Neuromuscular Stimulator

$355.00

SKU:
82/CHA77722
Quantity
  • Product Description

      Intelect® Portable Electrotherapy Neuromuscular Electrical Stimulation. For the prevention of disuse atrophy as well as muscle re-education, increasing range of motion and increasing circulation. Includes leadwires, 9-volt battery, Dura-Stick® Self-Adhesive Electrodes and a convenient carrying case.

      Features & Benefits - Dual Independent Channels - Adjustable Pulse Amplitude, Pulse Frequency and Contraction/Relaxation - Time parameter

      Indications
      This NMES is intended to be used in:
      1. Relaxation of muscle spasm.
      2. Prevention or retardation of disuses atrophy.
      3. Increase local blood circulation.
      4. Muscle re-education.
      5. Immediate post surgical simulation of calf muscles to prevent venous thrombosis.
      6. Maintaining or increasing range of motion.

      Channel

      Dual, isolated

      Waveform

      Asymmetric biphasic square pulse

      Pulse Amplitude

      0-80 mA each channel adjustable (500 ohm load)

      Pulse Width

      Fixed 250æ sec

      Pulse Frequency

      5, 30, 100 Hz

      Dimensions

      1" x 2.4" x 3.6" (26 x 62 x 91 mm)

      Weight

      4.4 oz (126 g)

      Safety Tests

      EN 60601-1, EN 60601-2, EN 60601-2-10


      Mode Interpretations
      "S" The pulses of CHANNEL 1 and CHANNEL 2 are synchronous.
      While Channel 1 is activated, Channel 2 will be activated simultaneously. The pulses active and inactive duration is controlled by ON TIME and OFF TIME.
      In this mode, ON TIME value should be set for more than ramp up value plus ramp down value. For example, if you set on ramp value for 2 seconds, ON TIME should be set for more than 4 seconds (2+2), i.e. ON TIME ≥ Ramp up + Ramp down
      "C" Both CHANNEL 1 and CHANNEL 2 are continuous pulses under an adjustable pulse rate and pulse width. The functions of ON TIME, OFF TIME, and RAMP cannot be set
      "A" The pulses of CHANNEL 1 and CHANNEL 2 are Alternative.
      When Channel 1 is activated, Channel 2 will be inactivated and vice versa.
      In this mode, ON TIME value should be set for more than ramp up value plus ramp down value; OFF TIME value should be set for more than ON TIME value. For example, if you set on ramp value for 2 seconds, ON TIME should be set for more than 4 seconds (2+2), and OFF TIME should be set for more than 4 seconds, i.e.
      ON TIME ≥ Ramp up + Ramp down